HOME    Cell Thawing System    GCCT-COP
GCCT-COP

GCCT-COP

COP/COC Vial Automated Water-Free Cell Thawing System

 

Vaccines, biologics, and other pharmaceuticals are typically stored and packaged in COP/COC vials. To address the needs of users who work with COP/COC vials, our company has developed the GCCT-Vial Automated Water-Free Cell Thawing System.

 

img1

 

The Disadvantages of Water Bath Thawing

The most common method for rapidly thawing frozen cell samples is to partially submerge cryovials in a 37 ℃ water bath for a few minutes and then observe the sample to determine if thawing is complete. However, this method has some significant drawbacks.

  • High risk of contamination: Biological samples in the cryovial have a very high chance of becoming contaminated.
  • Not a sterile process: A water bath cannot be used as part of a sterile process.
  • Subjective and inconsistent: Different operators may have significant variations in determining the thawing time, final temperature, and endpoint, as the process relies on experience.
  • Regulatory limitations: The use of water baths is restricted in GMP or clinical settings.
  • Non-standardized: The process is not standardized or intelligent, making it impossible to integrate into automated cell culture equipment.

 

 

img2

 

Introducing the GCCT-Vial Automated Water-Free Cell Thawing System

GCCT-Vial system helps to eliminate the potential contamination risks associated with water bath thawing for biological samples in vials. When used with the GMP software -Automated Cell Thawing Management SystemGCCT-MSenables data management and traceability, standardizing the thawing process. 

 

img3

 

Product Features

  • Safe: The system eliminates the high risk of contamination from water baths, ensuring sample safety.
  • User-friendly: It is easy to operate, with controls available through both software and physical buttons. 
  • Standardized: With built-in control procedures, the system standardizes the thawing process, removing the need for subjective judgments.
  • Intelligent: Users can set their own thawing conditions and integrate the instrument into automated cell culture equipment, such as isolators.
  • Real-time tracking: The system records the thawing temperature in real-time and displays the temperature curve and data.
  • Traceable data: The software automatically saves the temperature curve and data, meeting GMP requirements.
  • Wide applicability: It is suitable for COP/COC cryovials stored under various conditions, including in liquid nitrogen, dry ice, and ultra-low temperature refrigerators.

 

img4

 

 

Specifications

 

 

Feature

GCCT-COP

Model

GCCT-Vial

Thawing Temperature

37-60℃

COP/COC Vial Volume

2-30ml (Can be customized for different brands and specifications)

Indicating Signal

LED lamp strip and buzzer

Temperature Acquisition Frequency

1Hz

Temperature Sensor

IR sensor

Temperature Accuracy

±0.5℃

Sample Protection

Emergency stop button; heating module actively separates from the vial to prevent overheating

Extensibility

Can be integrated into automated cell culture equipment

Power

AC100-240V, DC 24V, 3.75A

 

Application Fields

  • Cell banks
  • Vaccines
  • Biologics
  • Powder injections
  • Freeze-drying

 

Ordering Information

  • GCCT-Vial: For 2-30ml vials (can be customized for different brands and specifications)
  • GCCT-MS: GMP software--Automated cell thawing management system

Grocellbio:Innovating the Future of Cell Thawing 

 